• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Synthes GmbH RETENTION PIN FOR SCREWDRIVER SHORT / XL25

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Synthes GmbH RETENTION PIN FOR SCREWDRIVER SHORT / XL25 Back to Search Results
Catalog Number 03.045.004
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/23/1982
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the tip of the retention pin was found missing during tray setup.There was no surgical delay.It was noted during manufacturers investigation that the tip was broken off.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11.Additional manufacturer narrative: h3, h6: per franchise complaint product investigation procedure for complaints for which a non-manufacturing related probable cause has been identified no manufacturing record evaluation is require.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned device revealed that the tip of the retention pin for scrdriver short / xl25 was broken off.The fragment was not received and no other issues were identified.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces like overtightening the device while assemblance; or by assembling the device in a misaligned position.Properly handling and¿attention to the approved use of the device diminishes the risk of failure.A dimensional inspection for the retention pin for scrdriver short / xl25 was not performed due to post manufacturing damage.A functional test was not performed as it is not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the retention pin for scrdriver short / xl25 would contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RETENTION PIN FOR SCREWDRIVER SHORT / XL25
Common Device Name
Screwdriver
Manufacturer (Section D)
Synthes GmbH
luzernstrasse 19-21
zuchwil 4528
SZ  4528
Manufacturer (Section G)
Monument
1101 synthes avenue
monument CO 80132
Manufacturer Contact
elba bello
1302 wright lane east
west chester, PA 19380
+1(908)386-3534
MDR Report Key50841297
Report Number8030965-2026-09591
Device Sequence Number18651885
Product Code HXX
UDI-Device Identifier10886982292513
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
PMA/510(K) Number
Exempt
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.045.004
Device Lot Number5135P66
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/02/2026
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 09/07/2026
Initial Report FDA Received Date09/21/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/27/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-