• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. HD CAMERA HEAD; AUTOCLAVABLE CAMERA HEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. HD CAMERA HEAD; AUTOCLAVABLE CAMERA HEAD Back to Search Results
Model Number CH-S190-08-LB
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported the autoclavable camera head had a separate issue.During the initial evaluation of the device it was observed, failed leak test.No patient involvement.
 
Additional Manufacturer Narrative
Based on the final investigation of the device, the initial evaluation findings were confirmed.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction failed leak test due to puncture on cable was traced to component failure.Date of event is unknown.Additional information will be provided if available.Olympus will continue to monitor the field performance of this device.Should any additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HD CAMERA HEAD
Common Device Name
AUTOCLAVABLE CAMERA HEAD
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
+1(508)207-7661
MDR Report Key50841339
Report Number3002808148-2026-33045
Device Sequence Number16018069
Product Code FET
UDI-Device Identifier04953170336669
UDI-Public04953170336669
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K102059
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberCH-S190-08-LB
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/26/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/27/2026
Initial Report FDA Received Date09/21/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-