| Model Number |
BF-H190 |
| Medical Device Problem Codes |
Device Reprocessing Problem (1091); Temperature Problem (3022)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/26/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported the bronchovideoscope exhibited a foreign object inside the scope.The issue occurred during reprocessing.There was no patient involvement.The device was returned to olympus for inspection.During the device analysis, a burnt distal end plastic cover and foreign material inside the channel was found.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|