| Model Number |
72404273 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/27/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient with an inflatable penile prosthesis (ipp) presented with a nonfunctioning device.A revision surgery was performed, during which the physician confirmed that a crack was found in the cylinder connecting tube, so the device was replaced.There were no patient complications.
|
| |
|
Additional Manufacturer Narrative
|
|
Model number/catalog number: 72404310.Serial number: n/a.Batch/lot number: 1000071350.Model/catalog description: pump ms.Unique identifier (udi) #: (b)(4).Model number/catalog number: 720182-01.Serial number: n/a.Batch/lot number: 1000061574.Model/catalog description: reservoir flat 100 ml.Unique identifier (udi) #: ((b)(4).The device is not available for analysis; therefore, no physical analysis of the product could be performed.The reported device performance allegation cannot be confirmed.Mechanical issue and hole/perforated are acknowledged within the instructions for use (ifu) and are not related to off-label use or failure to follow instructions.
|
| |
|