| Model Number |
LifeVest WCD 4000 System |
| Medical Device Problem Code |
Structural Problem (2506)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/18/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A us distributor reported that an electrode belt had non-functional ecgs.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation of electrode belt has been completed.The reported problem (non-functional ecgs) has been confirmed.Upon investigation the electrode belt failed an ecg fall-off test.This failure was due to the j704 connector broken from the distribution node pca.The root cause for the broken wires could not be positively identified.There were no adverse events associated with the damaged belt.
|
| |
|