| Model Number |
LifeVest WCD 4000 System |
| Medical Device Problem Code |
Structural Problem (2506)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/15/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A us distributor contacted zoll to report that a patient was experiencing "add gel/replace belt" messages.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation of electrode belt has been completed.The reported problem (add gel / replace belt messages) has been confirmed.Upon investigation the electrode belt did not pass essential performance testing.The cause for the failure was isolated to a pulse wire broken from the rear 1 wire therapy electrode.The root cause for the open wire was unable to be positively identified.There were no adverse events associated with the damaged belt.
|
| |
|