| Model Number |
CF-H190L |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer returned the colonovideoscope, which is an olympus asset with no reported complaint.During the evaluation of the device, burn mark at light guide lens was observed.The service repair technician indicated that the most likely cause of the burn mark at light guide lens was due to component failure.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of event is unknown.Additional information will be provided if available.The device was returned to olympus for inspection.A definitive root cause could not be identified.Based on the results of the investigation, the most probable cause of the complaint was traced to component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|