MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
|
Back to Search Results |
|
| Model Number |
8637-40 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/26/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Information was received from multiple sources (manufacturer representative, healthcare provider) regarding a patient who was receiv ing dilaudid (hydromorphone) (15 mg/ml at 0.104 mg/day) and bupivacaine (10 mg/ml at 0.069 mg/day) via an implantable pump for spinal pain indications.It was reported that a segment of the patient¿s previous 8780 catheter was cut, removed and discarded during a prior spinal fusion procedure.The length of the discarded segment is unknown.During the current catheter revision, the physician elected to replace only the spinal segment and estimated the length of the remaining catheter based on available measurements rather than replacing the entire catheter as recommended.The calculated catheter length of 108.8 cm was therefore not definitively confirmed.As a result, the exact length of the remaining catheter could not be established during the current revision. the patient was initially placed on minimum rate because the programmed drug concentration would otherwise have resulted in a higher initial dose.The new catheter was not primed because the patient had not been receiving medication prior to the revision.During the catheter revision procedure, a dye study was performed under fluoroscopy and confirmed that the remaining intrathecal catheter was functioning.The issue is resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
Continuation of d10: product id 8780 serial# (b)(6)implanted: (b)(6)2020 product type catheter section d information references the main component of the system.Other relevant device(s) are: product id: 8780, serial/lot #: (b)(6), ubd: 02-nov-2022, udi#: (b)(4).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|