It was reported that during the catheter exchange, health professional confirmed that approximately 8 ml of retained fluid had been drained and attempted to remove the foley catheter; however, it did not fall out.A computed tomography (ct) scan revealed that the balloon section contained no retained fluid but appeared to be filled with air.Therefore, following the troubleshooting protocol, health professional removed the catheter using the non-rupture method.It was also mentioned that health professional was present for the tube removal.There is residual air in the balloon section of the device.Based on the ct images taken prior to tube removal, it appears there was slightly more residual air at that time.Need to verify this.Since this patient was transferred from another hospital, the product code is believed to be 7760014lr.
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|