| Model Number |
H74939206100870 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Failure to Deflate (4060)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/21/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the balloon could not be withdrawn requiring additional intervention.A 10.0 x80, 75cm charger balloon catheter was advanced for dilatation.During the venous balloon dilatation procedure, the first inflation was completed successfully.However, during the second inflation attempt, the balloon could not be withdrawn.Attempts to advance the device further were also unsuccessful.The balloon was pulled back along the vessel to the sheath entry site.The device was then removed through the sheath, and a needle was used to puncture the balloon to facilitate complete extraction.The procedure was completed with another of same device.There were no patient complications as a result of this event.The patient was stable post procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
D2b: pro code: (product code): fge, lit.E1: initial reporter phone: (b)(6).G4: premarket / 510(k): k112697, k112701, k141521.Good faith effort attempts were made to try and retrieve the device status and additional details regarding the reported event, but further information was unable to be obtained.
|
| |
|