| Model Number |
G56572 |
| Medical Device Problem Code |
Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/26/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During an endoscopic hemostasis procedure in the stomach, the physician used a cook hemospray endoscopic hemostat.It was reported that the powder was not coming out.The tool "[catheter]" was already in the patient "[unable to spray - subject of report]".A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
| |
|
Additional Manufacturer Narrative
|
|
Continued:
section e: phone: (b)(6).Investigation evaluation:
a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion:
we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.A corrective action (capa) was initiated to further investigate device failure due to being unable to spray powder.This device is within the scope of the capa.Prior to distribution, all hemospray endoscopic hemostats are subjected to a visual inspection to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action:
a corrective action (capa) was initiated to further investigate device failure due to being unable to spray powder.The product said to be involved is included in the scope of the corrective action.A review of the complaint history was conducted.Quality assurance will continue to monitor for complaint trends.
|
| |
|
Search Alerts/Recalls
|