| Model Number |
35700BAX2 |
| Medical Device Problem Code |
False Alarm (1013)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a spectrum pump alarmed downstream occlusion during programming/setup.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was received for evaluation.During functional testing, the customer reported ¿downstream occlusion¿ was unable to be reproduced.The device was unable to complete testing for downstream occlusion due to service event of system error 321 after loading tube.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.The reported condition was verified.The cause of the condition was determined to be the exposed force sensor thermistor wire.The force sensor will be replaced to address this issue.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|