The customer reported suction doesn't stick and stated that the suction handle slides right out; the same thing happened when 4 more within that lot were sampled.Medline quality received 1 ea, and functional inspection was conducted with the provided sample; no issues were found, the yankuuer handle used in the test was provided by medline, the dhr was checked and no anomalies were recorded, and the customer did not respond when asked what handle they were using.At this time, the reported concern could not be confirmed, as no issues were found.The root cause cannot be identified, and the division will continue to monitor this issue for trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|