The account reported finding "hub broken during suctioning." a sample was received and observed to be broken, confirming the complaint; the detachment was at the position between the s-shaped main body and the negative pressure connector.Device history record, incoming inspection, and shipping records were reviewed and found no similar abnormalities.The root cause could not be determined.No trends were found for this item, and in the event additional information becomes available this complaint will be reopened, updated, reevaluated, and investigated, as necessary.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.This mdr is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|