| Model Number |
G58010 |
| Medical Device Problem Code |
Failure to Fire (2610)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During an endoscopic clipping procedure, the physician used two cook instinct plus endoscopic clipping devices.It was reported that the clips did not deploy.They get on tissue and they will not fire but they were able to reopen the clips to remove them from tissue.Another clip was grabbed to complete the procedure.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A follow-up emdr will be submitted within 30 days of receipt of new information.
|
| |
|