• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; STARTER KIT, GLUCOSE, EVENCARE G2 BT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; STARTER KIT, GLUCOSE, EVENCARE G2 BT Back to Search Results
Catalog Number MPH1545BT
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/16/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the glucose monitor is not working properly and believes it is defective.The device is not reading accurately or giving accurate readings.
 
Additional Manufacturer Narrative
A complaint was received, and the customer reported receiving a medline glucose starter kit that has inaccurate readings.At this time, a sample(s) was not received for investigation; however, investigation included review of recent complaint trends for the last three trend reports and the vendor was notified.The complaint was not confirmed due to the fact that this issue can only be confirmed by further investigating the sample or a provided photo, and no trends were identified for the specified material number and its corresponding defect type.Cause was documented as sample required to determine cause; based on the review of the details provided for the complaint investigation, medline was unable to confirm the complaint issue or that the product failed to meet specifications.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
STARTER KIT, GLUCOSE, EVENCARE G2 BT
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
narendra nambula
three lakes drive
northfield, IL 60093--275
MDR Report Key50842615
Report Number1417592-2026-46583
Device Sequence Number10039783
Product Code NBW
UDI-Device Identifier00197344007595
UDI-Public00197344007595
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K212140
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMPH1545BT
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/16/2025
Initial Report FDA Received Date09/21/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-