| Model Number |
12773-03 |
| Medical Device Problem Code |
Malposition of Device (2616)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/31/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported this was a venotomy closure of a right common femoral vein using a prostyle device after a cardiac ablation procedure with a small-bore sheath.Reportedly, after deploying the needles, it was observed the suture was caught around the soft tissue of the tissue tract.The knot was visualized in the tissue tract and was unable to move down to the vein.Therefore, the device was removed and replaced with another prostyle device.However, the same issue occurred.Manual compression was then used to achieve hemostasis.There were no adverse patient effects.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed there is no indication of a lot specific issue.Based on the information received, the investigation determined that the reported issue appears to be related to operational context.In this case, it is likely that interaction with the patient anatomy or friable femoral vessel contributed to the reported difficulty.Based on the results of the complaint investigation, there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.The additional device referenced in b5 is being filed under a separate medwatch report.
|
| |
|
Search Alerts/Recalls
|
|
|