Nalu Medical, Inc Nalu Neurostimulation System; Peripheral Nerve Stimulator Back to Search Results
Model Number 72003
Medical Device Problem Code High impedance (1291)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/23/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during trial procedure, both leads were showing red ohms.Both leads were wiped multiple times, the electrode interfacing cable (eic) and eic extension were swapped and issue persisted.New leads were opened, and the red ohms cleared.
 
Additional Manufacturer Narrative
After conducting a post-acquisition, retrospective review of nalu peripheral nerve stimulation complaint records, boston scientific determined that this event meets mdr reporting criteria.This incident is being re-opened/updated per pcar42 remediation activities.No product was able to be salvaged from this case; therefore, the cause of the reported symptoms could not be determined based on the available information.Mdr no.3015425075-2026-00730 will address lead udi: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Nalu Neurostimulation System
Common Device Name
Peripheral Nerve Stimulator
Manufacturer (Section D)
Nalu Medical, Inc
2320 faraday ave
suite 100
carlsbad CA 92008
Manufacturer (Section G)
Nalu Medical, Inc
2320 faraday ave
suite 100
carlsbad CA 92008 -724
Manufacturer Contact
rebecca hess
2320 faraday ave
suite 100
carlsbad, CA 92008--724
MDR Report Key50842915
Report Number3015425075-2026-00780
Device Sequence Number18757315
Product Code GZF
UDI-Device Identifier00812537030555
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K183579
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number72003
Device Lot Number9459460
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/23/2025
Initial Report FDA Received Date09/21/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/29/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-