| Model Number |
72404131 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
09/11/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Correction to g2: report source\
\
model number/catalog number: 72400024\
serial number: n/a \
batch/lot number: 1100634467\
model/catalog description: balloon fgs 61-70 cm\
unique identifier (udi) #: ((b)(4) \
\
model number/catalog number: 72404127\
serial number: n/a \
batch/lot number: 1100710614\
model/catalog description: control pump fgs iz\
unique identifier (udi) #: (b)(4) \
\
the device is not available for analysis; therefore, no physical analysis of the product could be performed.The reported device performance allegation cannot be confirmed.The device history record (dhr) review confirmed that the device met all material, assembly and performance specifications prior to distribution.A review of the device instructions for use (ifu) was completed and did not reveal any evidence of device misuse, off label use, or failure to follow instructions.A risk review was completed and confirmed that the event of pain and mechanical issue was defined in the product risk documentation.This event type has been accounted for during analysis to support acceptable risk benefit for the product.Based on the information available the cause that contributed to the reported event cannot be established.Therefore, a conclusion code of cause not established was assigned to this investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
Concomitant products implanted at the time of the event:\
model number/catalog number: 72400024.\
serial number: (b)(6).\
batch/lot number: 1100634467.\
model/catalog description: balloon fgs 61-70 cm.\
unique identifier (udi) #: (b)(4).\
\
model number/catalog number: 72404127.\
serial number: n/a.\
batch/lot number: 1100710614.\
model/catalog description: control pump fgs iz.\
unique identifier (udi) #: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with an artificial urinary sphincter (aus) stated his device stopped functioning following implantation.The patient also stated he experienced pain.A surgical procedure was performed to remove the aus.No additional information was provided.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with an artificial urinary sphincter (aus) stated his device stopped functioning following implantation.The patient also stated he experienced pain.A surgical procedure was performed to remove the aus.No additional information was provided.
|
| |
|
Search Alerts/Recalls
|