It was reported that during a laparoscopic cholecystectomy procedure, the surgeon attempted to deploy clips using the device; however, no clips were loaded or deployed despite multiple firing attempts.The issue resulted in a delay to the procedure.No patient consequences.
|
|
(b)(4)
date sent 9/21/2026
d4 batch # a70180.Investigation summary
the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the el5ml device was returned with no apparent damage, and the opened packaging was returned along with the instrument.In an attempt to replicate the reported event, the device was tested for functionality.During testing, the device was fired; however, the clips failed to advance into the jaws.Further examination revealed that the tip of the advancer was bent.The instrument was subsequently disassembled, and the bent advancer condition was confirmed.Additionally, nine (9) clips were found remaining within the clip track at the time of disassembly.The event reported was confirmed and it is related to improper use of the device.Prior to loading a clip in the jaws, ensure that the demarcation between the jaws and the device shaft is past the end of the trocar cannula.Additionally, excessively applying a side load to the jaws, causing them to partially collapse, could result in a clip malformation.The device jaws should be fully open and parallel upon initiating the firing of the device.As part of ees quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch a70180 number, and no non-conformances were identified.
|