MAUDE Adverse Event Report: INTUITIVE SURGICAL, INC None; ENDOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL, INC None; ENDOSCOPE Back to Search Results
Model Number 470066-13
Medical Device Problem Code Image Orientation Incorrect (1305)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/28/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a da vinci-assisted surgical procedure, while targeting the customer found that the image angle shown by the 0 degree endoscope did not match the horizon indicator shown on the system.The customer reseated the endoscope and the issue was resolved.The procedure was completed with no reported injury.Isi followed up with the initial reporter and obtained the following additional information: the customer did not observe any damage or missing screws on the endoscope base.The endoscope was not manually rotated prior to the issue.The endoscope will be returned.
 
Additional Manufacturer Narrative
An investigation was completed to determine the cause of this reported event.The reported event was addressed with phone support.The customer reported a mismatch between the camera horizon indicator and actual camera orientation during targeting.The technical support engineer (tse) recommended to reattach the camera which resolved the issue.No site visit was conducted.The system was working properly, and no additional action was required as the issue was resolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
None
Common Device Name
ENDOSCOPE
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 95051
+1(408)523-2100
MDR Report Key50843379
Report Number2955842-2026-39442
Device Sequence Number9957631
Product Code GCJ
UDI-Device Identifier00886874120293
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K232610
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 09/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number470066-13
Device Catalogue Number470066
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/28/2026
Initial Report FDA Received Date09/21/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/24/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
da Vinci Instruments and Accessories
Patient SexUnknown
-
-