| Model Number |
LifeVest WCD 4000 System |
| Medical Device Problem Code |
Structural Problem (2506)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/17/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A us distributor reported that an electrode belt had non-functional therapy electrodes.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation of electrode has been completed.The reported problem (non-functional te) has been confirmed.Upon investigation the electrode belt did not pass a therapy electrode recognition test.The cause for the failure was isolated to an open pulse wire in the cable connecting the distribution node (dn) and ecg "b".The root cause for the open wire could not be positively identified.There were no adverse events associated with the damaged belt.
|
| |
|