| Catalog Number |
242268 |
| Medical Device Problem Codes |
Break (1069); Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that prior to a shoulder arthroscopy surgical procedure it was observed that the 4k c-mnt scp,4.0,30,167,mitek scope device was cracked.It was reported that a new scope was attached and the device worked fine.During in-house engineering evaluation it was determined that the device was broken (2+ pieces): chipped/shaved/delaminated.It was reported that there was no patient injury.The exact date of the event was not reported; however, it was reported that the event occurred in 2026.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative:
investigation summary = > the actual device and a photo were returned for evaluation.Visual inspection of the returned photo found no observations pertaining to the nature of the reported event or any other anomaly.The product was returned to depuy synthes service center for evaluation then, the device was forwarded to oem and the following defects were identified:
- other
damages caused by customer
- outer tube damaged, distal tip
distal tip damaged deep scratches/knicks
shaver damage to distal tip
fibers sunken in
- illumination
distal tip fiber damaged
- optical system, optical components
broken lenses in optical system
distal cover glass/negative damaged
- cosmetic issue
scratches/dents
rust/discoloration
- engraving/identification
datamatrix code level a
visual : scratched/nicked, broken (2+ pieces) : chipped/shaved/delaminated, visual : deformed/bent and labeling : illegible etch
the defective parts needs to be replaced to resolve the issues.The reported complaint was confirmed based on the service evaluation results.The investigation traced the potential cause to user error.It has been determined that no corrective or preventive actions are proposed.There is no indication that a design or manufacturing issue caused the reported condition.As part of depuy synthes quality processes, all devices are manufactured, inspected, and released in accordance with approved specifications.Additional monitoring for potential safety signals is conducted through complaint trend analysis as part of post-market safety surveillance activities.Device history lot = > a manufacturing record evaluation was performed for the finished device serial number, and no non-conformance was identified.
|
| |
|