| Model Number |
M001171030 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/19/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Reportable based on device analysis completed on (b)(6)2026.It was reported that balloon expansion issue occurred.The target lesion was a tight stenotic occlusion located in the iliac artery.A 7/2/100 occlusion balloon catheter was advanced for dilation.During the procedure, the balloon did not fully expand.The procedure was completed with a different device.There were no patient complications as a result of this event.However, device analysis revealed that the balloon ruptured.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results:
device evaluated by mfr: the occlusion balloons device was returned to our post market quality assurance laboratory.Visual and microscopic inspection was performed and identified that the balloon has ruptured.This concludes the product analysis.Device history record (dhr) review:
the device history record review confirmed that the device met all material, assembly, and performance specifications prior to distribution.Investigation conclusion:
based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of adverse event related to procedure.
|
| |
|