The customer reported a low air loss mattress was "triggering the service light" and stated, "i've tried it on multiple mattresses and it's doing the same thing." a sample of the pump only was provided, attached to a mattress for evaluation, and it was confirmed that the pump would not fully inflate the mattress; the returned pump was not compatible with the mattress system associated with the order number provided.The reported problem was confirmed during functional testing, and the cause was documented as "cannot determine cause w/received sample." there was no reported death or serious injury associated with this event, and the product was not used for treatment or diagnosis.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|