| Model Number |
2110 |
| Medical Device Problem Code |
Pumping Problem (3016)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/25/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during testing pump giving error 41622 after the first pump sequence.The event occurred during testing.During the investigation, it was confirmed that the high-priority error 41622 "error_motor_timeout" was displayed and recorded in the event log of the returned pump.This failure occurred during testing; however, if it were to recur during infusion, it could result in an interruption of therapy.There are no patient involvement and adverse event reported. .
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect pump was returned for evaluation, and the event history log (ehl) was reviewed.An error code 41622 was identified in the event log.A review of the 12-month service history found no prior service records.Visual and functional inspections revealed a contaminated uso seal, and the device failed the motor test, downstream occlusion test, upstream occlusion sensor test, and battery fall-out test.The reported issue was confirmed during the evaluation, and the root cause was determined to be a defective flex sensor.Corrective action includes replacement of the uso seal and cam flex sensor.
|
| |
|