• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; LIFT,PATIENT,STAND ASSIST,ELECTRC,40

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; LIFT,PATIENT,STAND ASSIST,ELECTRC,40 Back to Search Results
Catalog Number MDS400SAB
Medical Device Problem Code Failure to Run on Battery (1466)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/12/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that lifts will not go up or down at all; all electrical components are fine and seem to be functional and in acceptable condition, but the lifts will not raise or lower.
 
Additional Manufacturer Narrative
The customer reported that the lifts have no function and will not raise or lower; when they tried to use the lift without the charger plugged in it would show a battery blinking and then shut off.Medline reached out to the customer to troubleshoot or to obtain additional information; the customer unplugged everything and re-plugged securely, swapped batteries and pendants, and the vendor contacted the customer.Investigation results: not confirmed.Cause was documented as ¿sample required to determine cause¿ and ¿unconfirmed defect,¿ replacement battery packs were issued to the customer, and the division will monitor through quarterly trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
LIFT,PATIENT,STAND ASSIST,ELECTRC,40
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
shivani gheewala
three lakes drive
northfield, IL 60093
MDR Report Key50844541
Report Number1417592-2026-46777
Device Sequence Number10050455
Product Code FSA
UDI-Device Identifier10193489019254
UDI-Public10193489019254
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMDS400SAB
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/12/2024
Initial Report FDA Received Date09/21/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-