| Model Number |
9560524 |
| Medical Device Problem Code |
Mechanics Altered (2984)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient having microendoscopic discectomy.It was reported that the fixing screw was rotated, it could not be fixed to the post, and handle does not rotate.Also the camera part at the tip is reported to be loose.There was no patient symptom reported.The patient was present at the time of the event, but since it was discovered during procedure preparation, there was no impact on the patient.There were no further complications reported regarding the event.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product analysis - product: 9560524, lot no: my23e002 during the incoming inspection, deformation of the threads of the handle screw and looseness of the holder were observed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|