| Model Number |
CDS0801-XT |
| Medical Device Problem Codes |
Difficult or Delayed Positioning (1157); Entrapment of Device (1212)
|
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); Unspecified Tissue Injury (4559)
|
| Date of Event |
08/28/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported on (b)(6) 2026 that the patient presented with grade 4 degenerative mitral regurgitation (mr) and small flail for a mitraclip procedure.During the procedure, mitraclip nt anterior gripper was unable to lower during grasping of the leaflets.Difficulty was encountered while capturing the leaflets due to sub-optimal imaging.Troubleshooting was performed and was unsuccessful.Multiple grasping and capturing attempts contributed to leaflet damage.The mitraclip was removed from the anatomy.Mitraclip xt was inserted into the anatomy.Additional chordal rupture was observed in the grasping area due to the chordal entanglement.Troubleshooting was performed and was unsuccessful in disengaging the clip from chordae.As a result, the mitraclip xt was implanted on anterior leaflet and chordae.The mr remained unchanged post procedure.There was no clinically significant delay reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device remains implanted in patient.The device will not be returned for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|