| Model Number |
12773-03 |
| Medical Device Problem Code |
Malposition of Device (2616)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/05/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported this was an arteriotomy closure of a right common femoral artery using a prostyle device after a percutaneous coronary intervention (pci) procedure with a 6f sheath.Reportedly, after deployment of a prostyle device, the formed knot came out of the anatomy with the device.It was suspected the suture was deployed in the tissue tract.A non-abbott device was then used to achieve hemostasis.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed there is no indication of a lot specific issue.Based on the information received, the investigation determined that the reported issue appears to be related to operational context.In this case, it is likely that interaction with the patient anatomy or friable femoral vessel contributed to the reported malposition of the suture.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
| |
|