| Catalog Number |
07P70-30 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/07/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported falsely elevated alinity i free t4 results for multiple patients.The following data were provided:
sid (b)(6)(female, born 1947)
(7sep2026)
initial: >64.35 pmol/l,
repeat: 19.13 pmol/l, 17.70 pmol/l.Sid (b)(6)(male, born 1979)
((b)(6) 2026)
initial: >64.35 pmol/l.Repeat: 15.31 pmol/l.No impact to patient management was reported.
|
| |
|
Event or Problem Description
|
|
The customer reported falsely elevated alinity i free t4 results for multiple patients.The following data were provided:
sid (b)(6) (female, born 1947)
(b)(6) 2026)
initial: >64.35 pmol/l
repeat: 19.13 pmol/l, 17.70 pmol/l.Sid (b)(4) (male, born 1979)
(b)(6) 2026)
initial: >64.35 pmol/l
repeat: 15.31 pmol/l.No impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.Section a patient information: refer to section b5 for complete patient information.
|
| |
|
Additional Manufacturer Narrative
|
|
Data and information provided by the customer were reviewed and support the complaint issue.A review of tracking and trending identified an increase in complaint activity for the alinity i free t4; however, no trends were identified for lot 87264ud00.A labeling review concludes that the complaint issue is adequately addressed.A device history record review did not identify any nonconformances, potential nonconformances or deviations associated with lot 87264ud00 and the complaint issue.Customer field data was used to assess the performance of the alinity i free t4 using worldwide data.The median patient result for the complaint lot is within the established limits, indicating that the reagent lot is performing acceptably in the field.Based on our investigation, no systemic issue or deficiency of the alinity i free t4 lot 87264ud00 was identified.
|
| |
|