Model Number M635WU60350
Medical Device Problem Code Failure to Seal (4070)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/02/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported an implant did not seal.A left atrial appendage (laa) closure procedure was performed, and a 35mm watchman flx pro closure device was deployed.A 3 mm posterior/lateral peri-device leak at 0 degrees and a 2.5 mm posterior peri-device leak at 135 degrees were considered acceptable, and the device was released.The patient was discharged following the procedure.At a routine follow up, transesophageal echocardiography (tee) was performed, and it was noted the posterior peri-device leak had increased to 3.74 mm.As a result, the patient is unable to discontinue their prescribed blood thinner medication.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WATCHMAN FLX Pro
Common Device Name
system, appendage closure, left atrial
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
ballybrit business park
galway H91 Y 868
EI   H91 Y868
Manufacturer Contact
mary crawford
4100 hamline ave n
arden hills, MN 55112
+1(651)440-0979
MDR Report Key50846219
Report Number2124215-2026-50268
Device Sequence Number10278498
Product Code NGV
UDI-Device Identifier00191506004620
UDI-Public(01)00191506004620(17)281203(10)38111128
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P130013
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM635WU60350
Device Catalogue NumberM635WU60350
Device Lot Number0038111128
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/17/2026
Initial Report FDA Received Date09/21/2026
Date Device Manufactured12/04/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age78 YR
Patient SexMale
-
-