Medos International SÃ rl HD LAPSCP,EP,5.5,30,300; Laparoscope, General & Plastic Surgery
|
Back to Search Results |
|
| Catalog Number |
242633 |
| Medical Device Problem Codes |
Material Fragmentation (1261); Scratched Material (3020)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during in-house engineering evaluation, it was determined that the hd lapscp,ep,5.5,30,300 device was chipped/shaved/delaminate.It was noted by the reporter that the device was scratched.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2026.All available information has been disclosed.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11: additional narrative:
investigation summary the product was returned to depuy synthes service center for evaluation then, the device was forwarded to oem and the following defects were identified:
repair level 3 major repair
- other
damages caused by customer
- outer tube damaged, needle
outer tube bent
- outer tube damaged, distal tip
distal tip damaged deep scratches/knicks
keel loose/raised keel sunken
- illumination
distal tip fiber damaged
- optical system, optical components
broken lenses in optical system
distal cover glass/negative damaged
- cosmetic issue
scratches/dents
rust/discoloration
visual : scratched/nicked, broken (2+ pieces) : chipped/shaved/delaminated, visual : deformed/bent and visual : corrosion/rusting/pitting
the defective parts needs to be replaced to resolve the issues.The reported complaint was confirmed based on the service evaluation results.The investigation traced the potential cause to user error.It has been determined that no corrective or preventive actions are proposed.There is no indication that a design or manufacturing issue caused the reported condition.As part of depuy synthes quality processes, all devices are manufactured, inspected, and released in accordance with approved specifications.Additional monitoring for potential safety signals is conducted through complaint trend analysis as part of post-market safety surveillance activities.
|
| |
|
Search Alerts/Recalls
|
|
|