| Model Number |
WA50040A |
| Medical Device Problem Codes |
Break (1069); Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/28/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that the rigid video laparoscope displayed image noise during therapeutic laparoscopy procedure while the patient was under sedation.The intended procedure was completed using a backup device.During the initial evaluation of the device, the fiber bonding in the outer tube area at the distal end was damaged, and no image was displayed.No patient harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The final investigation has been completed.The device was returned to olympus for inspection, and the reported failure and failure identified during the inspection have been confirmed.A root cause could not be identified.Based on the results of the investigation image is not displayed due to video cable is broken also for the other malfunction based on the results of the investigation, the most probable cause of the complaint was traced to component failure.Olympus will continue to monitor the field performance of this device and evaluate any additional information received through post-market surveillance activities.
|
| |
|