The customer reported that the gloves got a hole in them while setting up a surgical table, and the glove was used for set up of a sterile procedural table with no biohazard involved, just contamination of sterile equipment.Inspection records for the reported sku lot were reviewed and the lot passed with no abnormalities noted; all scar and capa records for the glove division were reviewed and there have been no opened scars or capas for the reported sku in the last 24 months; trending was reviewed and there has been 1 additional reported complaint for this issue for this sku in the past 24 months.Photo received: no; sample received: no; picture / sample evaluation: n/a; in the absence of a physical sample and without high-quality, detailed photos, medline was unable to confirm the presence of a defect during this investigation.No root cause was identified, and the quality team will continue to monitor for trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|