Product was returned and evaluated.The reported issue of the unit having intermittent power was not confirmed.Functional testing was performed using both aa batteries and the ac power adapter.The unit remained powered on throughout the evaluation period, and the reported intermittent power condition could not be reproduced.Following testing, the unit was disassembled and the printed circuit board was visually inspected.No damaged components, discoloration, or other abnormalities were observed.Based on the evaluation conducted, no evidence was identified to confirm the reported intermittent power issue.Manufacturer reference file: (b)(4).
|