| Model Number |
HistoCore PELORIS 3 Automated Tissue Processor |
| Medical Device Problem Code |
Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On 24 august 2026, the customer contacted leica biosystems and reported that ¿tissue through cleaning cycle.Tissues were left in the processor and went through the cleaning cycle, are there any recommendations for ways to mitigate the damage? the blocks are quite crispy.¿
on 29 august 2026, the leica senior applications specialist documented ¿¿all 8 cases have been signed out, diagnosed.Customer also stated that if needed, they could go back to gross tissue for re-grossing for processing.¿.
|
| |
|
Additional Manufacturer Narrative
|
|
The root cause for the sub-optimal tissue processing reported by the customer was due to a user error.Specifically, the information indicates that a user failed to follow the manufacturer instructions detailed in section 3.2 of the leica histocore peloris 3 user manual, which contains the following specific warning: ¿do not use cleaning protocols for reprocessing as the dry step will damage the tissue.¿ accordingly, exposure of the patient tissue samples to the retort cleaning protocol would have adversely impacted the patient tissue samples.Although the information available indicates that all patient cases involved in this event were diagnosable, the circumstances involved in this complaint have previously resulted in serious injury.No adverse impact on either the quality of processing of patient tissue samples or to any patient(s) has been reported to the manufacturer in association with the circumstances involved in this complaint.Manufacturer evaluation of the information available showed that the instrument functioned as designed during execution of the affected tissue processing run.
|
| |
|
Search Alerts/Recalls
|