| Model Number |
8015 |
| Medical Device Problem Code |
Inaccurate Delivery (2339)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/30/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the patient was receiving a continuous bivalirudin infusion that was ordered to run at 0.158 mg/kg/hr.The infusion was started at the correct rate 0.158 mg/kg/h (12.6 ml/hr) via interoperability.It was later discovered running at the incorrect rate of 0.1593 mg/kg/hr (12.7 ml/hr) for approximately two hours starting at 5:41am.There was patient involvement and no harm.Additional information provided 02sep2026: per review of the event logs, it appears that the clinician inadvertently pressed the "up arrow" key prior to pressing start at 05:41:17.This increased the rate from 12.6 ml/hr to 12.7 ml/hr.
|
| |
|
Additional Manufacturer Narrative
|
|
Bd technical support troubleshoot with customer over the phone.The event logs have been received and the evaluation is pending.A follow up report will be submitted with investigation results once the evaluation has been completed.Per 803.52(f)(11)(iii) the information provided represents all of the known information at this time.The complainant or reporter was unable or unwilling to provide any further patient, product, or procedural details to the manufacturer.
|
| |
|