| Catalog Number |
D141201 |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure and after ablation with a varipulse¿ bi-directional catheter, they noticed the smartablate irrigation tubing set was disconnected at the catheter end (no damage noticed).There were no errors displayed.The pump was flowing, just not into the catheter, and there was no pool of fluid found.They also cardioverted, and there was no injury to patient.Additional information indicated that it is not known exactly how it became disconnected.Mid-way through the case, the physician noticed that it got disconnected and the issue was resolved.It is suspected that it came off during the cardioversion (planned), however, it remains unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
|
| |
|