| Model Number |
380620-47 |
| Medical Device Problem Code |
Visual Prompts will not Clear (2281)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/28/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that prior to the start of a da vinci-assisted surgical procedure that non-recoverable fault 32056 occurred against universal surgical manipulator (usm) 4.As system was onsite, the intuitive surgical inc technical support engineer (tse) reviewed the system logs and found faults against a board inside usm 4.The tse asked the caller if they would be able to proceed with three usms; the caller was unsure as the surgeon was not in the room yet.The procedure outcome not provided.The ports were not in place when the issue occurred.Isi contacted the initial reporter and obtained the following information: there was no harm to the patient, and the procedure was aborted to an open procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
An intuitive surgical, inc.(isi) field service engineer (fse) was dispatched to the customer site to further investigate the reported complaint.The fse was able to reproduce the reported complaint and replaced the universal surgical manipulator (usm).The system was tested and verified as ready for use.As of the date of this report, the usm has not yet been received by isi for evaluation.
|
| |
|