Model Number 380620-47
Medical Device Problem Code Visual Prompts will not Clear (2281)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/28/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that prior to the start of a da vinci-assisted surgical procedure that non-recoverable fault 32056 occurred against universal surgical manipulator (usm) 4.As system was onsite, the intuitive surgical inc technical support engineer (tse) reviewed the system logs and found faults against a board inside usm 4.The tse asked the caller if they would be able to proceed with three usms; the caller was unsure as the surgeon was not in the room yet.The procedure outcome not provided.The ports were not in place when the issue occurred.Isi contacted the initial reporter and obtained the following information: there was no harm to the patient, and the procedure was aborted to an open procedure.
 
Additional Manufacturer Narrative
An intuitive surgical, inc.(isi) field service engineer (fse) was dispatched to the customer site to further investigate the reported complaint.The fse was able to reproduce the reported complaint and replaced the universal surgical manipulator (usm).The system was tested and verified as ready for use.As of the date of this report, the usm has not yet been received by isi for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
daVinci X
Common Device Name
PATIENT SIDE CART, 4-ARM
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 95051
+1(408)523-2100
MDR Report Key50850381
Report Number2955842-2026-39635
Device Sequence Number21206772
Product Code NAY
UDI-Device Identifier00886874115404
UDI-Public(01)00886874115404
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K171294
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Nurse
Remedial Action Other
Type of Report Initial
Report Date (Section B) 09/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number380620-47
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 08/28/2026
Initial Report FDA Received Date09/22/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/17/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
da Vinci Instruments and Accessories.
Patient SexUnknown
-
-