| Model Number |
430151-02 |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
07/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted surgical procedure, the front part of the 8mm fenestrated bipolar forceps instrument did not fully engage.Upon removal and inspection, it was discovered that the instrument did not close completely even manually, so the arm was replaced with a new one and the surgery was performed.The procedure was completed with no reported injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The 8mm fenestrated bipolar forceps instrument was analyzed and found to have one severely bent grip, causing misalignment of the grip tips.There was a 2.24 mm offset at the tips.The grips were not found to have cracking damage.Additional observation related to the customer complaint: the instrument was found to have a dislodged molded insulator on the jaw.The grip tip for the grip with the dislodged molded insulator does appear to be bent.Components adjacent to the dislodged molded insulator do not show damage.The complaint was confirmed by failure analysis.The probable root cause of severely bent grips is attributed to damage either during use or reprocessing, as severe misalignment of grip tips can be due to accidental drops, incorrect instrument tray or sink sizes for reprocessing, or overloading the tips.
|
| |
|