Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The 8mm permanent cautery hook (pch) instrument was analyzed and found to have a broken distal clevis at the base of the clevis.The broken piece was returned, measuring roughly 0.208# x 0.449# in size.Clevis does show abrasions or scratches on the outer surface.The component is broken and there may be potential missing pieces, but the size of the missing pieces cannot be confirmed.Components adjacent to the broken clevis do not show damage.The complaint was confirmed by failure analysis.The probable root cause of the damaged conductor wire is attributed to component susceptibility to damage through external collisions, during use, or during reprocessing.Additionally, conductor wire management issues can result in damage to the wire itself, crimp, and insulation.
|