Medtronic Ireland HAWKONE 6FR; CATHETER, PERIPHERAL, ATHERECTOMY Back to Search Results
Catalog Number H1-M
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/20/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during atherectomy of a calcified, 100% stenosed target lesion in the proximal left posterior tibial artery with a vessel diameter of 5.0 mm, the hawkone h1-m device thumb switch would not move to the off position after cutting, including after removal from the body.Embolic protection was used with a spider 4.0 device, the vessel was pre-dilated with a nanocross 3 x 120 device, and post-dilated with an in.Pact 5 x 250 device.The hawkone h1-m device was safely removed from the patient.The procedure was completed using a new hawkone ls device.The event delayed surgery by 10 minutes.No patient complications have been reported as a result of this event.Additional information: when the device was returned for evaluation the cutter was detached/broken from the driveshaft.
 
Additional Manufacturer Narrative
Product analysis: the device was returned with the thumbswitch in the ¿on¿ position, and the cutter positioned in the cutter window.The thumbswitch was advanced but the thumbswitch and the cutter would not move.The housing was removed at the hinge pins, and a visual inspection found the that the cutter was detached/broken from the driveshaft.The detached cutter remains inside the housing.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HAWKONE 6FR
Common Device Name
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
Medtronic Ireland
parkmore business park west
galway
EI 
Manufacturer (Section G)
Medtronic Ireland
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
+353(091)708096
MDR Report Key50850904
Report Number9612164-2026-04752
Device Sequence Number9923308
Product Code MCW
UDI-Device Identifier00643169968332
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K161361
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberH1-M
Device Lot Number0013371731
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/18/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/21/2026
Initial Report FDA Received Date09/22/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/03/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-