A review of the complaint history for (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for (b)(6) was performed from the date of manufacture, 07-feb-2024 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the user was unable to issue medication.A technical support specialist (tss) remotely connected to the workstation and reproduced the reported issue with the customer.Investigation found that the lorazepam order was in requested rx verify status with a medication quantity of 0, preventing issuance.The customer was advised that the quantity must be greater than 0 and the order must be approved through rx verify before the medication can be issued.The customer was instructed to contact the pharmacy to update and approve the order.The customer also reported a similar issue involving an antibiotic medication.Further investigation revealed that the drawer zones were not properly configured, with only zone 6 populated.The drawer zones were repopulated, restoring functionality, and the customer successfully issued the medication.During the call, the customer was advised that the lorazepam order had been rejected by the pharmacy.The customer was advised to follow up with the pharmacy regarding the rejected order and customer acknowledged the recommendation.No downtime and no further problem was reported.The system functioned as intended after the technical support specialist troubleshot the device.
|