|
C.R. Bard, Inc. (BASD) -3006260740 Hickman CV Catheter, Triple-Lumen, 12.5F; Chronic Catheters
|
Back to Search Results |
|
| Catalog Number |
0600650 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/31/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2026, a patient underwent a central venous catheter placement procedure using hickman/leonard/broviac central venous catheter.After 15 days, the catheter had ruptured and was leaking.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: as the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, photos and a video sample were provided for review.The investigation of the reported event is currently underway.Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Search Alerts/Recalls
|
|
|