| Model Number |
500001004500 |
| Medical Device Problem Code |
Application Program Problem (2880)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/26/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that when using the bd pyxis¿ es refrigerator, refrigerator failed to recover.Customer tried to reboot and recover but issue wasn't resolve.The customer stated that this malfunction caused a delay in dispensing the medications to the patient. however, there were no adverse events or injuries reported based on this event.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 06-nov-2024 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that refrigerator door lock state became out of synchronization.The field service engineer (fse) powered down both the fridge and med station completely, then restarted the fridge followed by the med station to reinitialize communication and lock control functions.The system functioned as intended after the field service engineer (fse) resolved the issue.
|
| |
|