• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Maquet Cardiovascular LLC VasoView Hemopro 2; Electrosurgical, Cutting & Coagulation & Accessories

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Maquet Cardiovascular LLC VasoView Hemopro 2; Electrosurgical, Cutting & Coagulation & Accessories Back to Search Results
Model Number VH-4000
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/28/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Tw (b)(4) event site name exceeds character limitations: (b)(6).The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.
 
Event or Problem Description
The hospital reported that vasoview hemopro 2 wouldn't burn halfway through the case.The harvester switched out cable, adaptor and power supply and the problem persisted.The surgeon switched out kit and problem was resolved.There was procedural delay of 67 min.No patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VasoView Hemopro 2
Common Device Name
Electrosurgical, Cutting & Coagulation & Accessories
Manufacturer (Section D)
Maquet Cardiovascular LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer (Section G)
Maquet Cardiovascular LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 07470
MDR Report Key50851935
Report Number2242352-2026-0000821
Device Sequence Number10053493
Product Code GEI
UDI-Device Identifier00607567700406
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K101274
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse Practitioner
Type of Report Initial
Report Date (Section B) 09/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVH-4000
Device Catalogue NumberVH-4000
Device Lot Number3000559188
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/04/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/28/2026
Initial Report FDA Received Date09/22/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/07/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
No concomitant products were used,
Patient Age56 YR
Patient SexMale
Patient Weight156 LB
-
-