| Model Number |
WA50082A |
| Medical Device Problem Codes |
Communication or Transmission Problem (2896); Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/26/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer returned the rigid video laparoscope, which is an olympus asset with no reported complaint.During the evaluation of the device repair technician found visible foreign material at cover glass of the insertion section and scope communication error (e226).There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The final investigation has been completed.The failure identified during the inspection have been confirmed.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: the most probable cause of the malfunction "scope communication error (e226)" is traced to component failure and for the malfunction "visible foreign material at cover glass of the insertion section" is cause not established.Olympus will continue to monitor the field performance of this device and evaluate any additional information received through post-market surveillance activities.
|
| |
|