| Model Number |
470057-08 |
| Medical Device Problem Code |
Image Orientation Incorrect (1305)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/29/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted high anterior resection surgical procedure, the 30-degree endoscope plus base rotated stiffly.Sometimes the up-and-down inversion did not work.The procedure was completed with no reported injury.Isi followed up with the initial reporter and obtained the following additional information: customer alleged that endoscope image was inverted.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The 30-degree endoscope plus was analyzed and placed through friction test using a screening aid to manually rotate the adapter to detect for friction.The camera instrument adapter did not rotate properly and/or noises were heard during testing and confirmed as binding.The endoscope was tested and placed on a camera attenuation tester or equivalent to measure transmittance but failed functional testing due to the measured output power not meeting specification limits.The complaint was confirmed by failure analysis.The probable root cause is attributed to component susceptibility to increased friction.
|
| |
|