| Model Number |
VH-3010 |
| Medical Device Problem Code |
Failure to Deliver Energy (1211)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/31/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The hospital reported that the t.W.Power supply began malfunctioning during a cabg procedure.The issue was noticed during sputtering when delivering energy.The customer did not report any information regarding beeping tones.Exchanging the cables/adapters did not resolve the issue; however, using a different power supply did.Power setting at 3.The hemopro power supply was first put into distribution in 2015 the dates for the hemopro extension cable and adapter are unknown; therefore, their ages cannot be determined.The power supply was placed on the endoscopic tower during use.It was not located in close proximity to other power-generating devices or equipment that would be expected to generate a significant amount of electrical power or cause interference.Minor delay of 2-3 minutes.No patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Event site name exceeds character limitations: (b)(6).The device has not yet been returned to maquet cardiac surgery for evaluation.We are following up with the customer for the return of the device.A supplemental report will be submitted if the device is received.
|
| |
|